The following comparison chart details the current product line of Analysis Tech Event Detectors, which comprehensively span the range of connector test specifications for transient resistance monitoring. In addition to this list of standard products, custom systems are available to suit particular test needs. Analysis Tech also encourages requests for custom systems to service new test requirements.
| |
Number of
Channels |
Sense
Current
(mA) |
Minimum
Duration |
Threshold Resistance (ohms) |
| 32EHD |
32 NC |
10 – 100* |
1, 10, 50 nsec* |
0.5 - 20 |
| 32-106 |
32 NC |
1 – 100* |
0.1, 0.5, 1.0 usec* |
0.1 - 2000 |
| 32-105 |
32 NC |
1 – 100* |
0.1, 0.5, 1.0 usec* |
0.1 - 2000 |
64-105
|
64 NC
|
1 – 100*
|
0.1, 0.5, 1.0 usec*
|
0.1 - 2000 |
96-105
|
96 NC
|
1 – 100*
|
0.1, 0.5, 1.0 usec* |
0.1 - 2000 |
128-105
|
128 NC |
1 – 100*
|
0.1, 0.5, 1.0 usec* |
0.1 - 2000 |
| 32HLV |
32 NC
|
2 |
0.5 usec
|
0.01 - 2 |
| 64STD |
64NC |
1.2 |
0.2 usec |
100 - 5000 |
| 128STD |
128 NC
|
1.2 |
0.2 usec
|
100 - 5000 |
256STD
|
256 NC
|
1.2
|
0.2 usec
|
100 - 5000 |
| * - indicates other values are available |
Notes:
| 1. |
The 105 and 106 Series Event Detectors are electrically and functionally identical. All 105 Series instruments are housed in a chassis capable of accepting up to 128 channels, while the 106 Series instruments have a maximum of 32 channels. |
| 2. |
The 32HLV Amplifier is an accessory to the 106 and 106 Series Event Detectors, and cannot be used as a stand-alone product. |
| 3. |
All STD Series Event Detectors are housed in a chassis capable of accepting up to 256 channels. The 64STD and 128STD models can be upgraded at any time. |
|
|
|
|