Recirculating Chiller Requirements

 

The TIM Testers require a recirculating chiller for coolant flow with sufficient temperature stability and flow rate for optimal test results.

Recommended Chiller Specifications
Flow Rate: 2.0 gal/min (8 liter/min) positive displacement pump
Pressure Limit: 50 psi (280 kPa)
Temperature Control Stability: +/- 0.1°C @ min. 500W (TIM 1300) or min. 900W (TIM 1400)
Set Point Range (minimum): 10°C - 35°C

Any chiller that fits these technical requirements can be used although a chiller that is controllable via serial communication port is most convenient. WinTIM software uses the serial control link to turn the chiller ON/OFF and to change the chiller setpoints. When using a chiller that is not equipped with serial control capability, WinTIM displays screen-prompts for the operator to perform these ON/OFF and setpoint changes manually. Although WinTIM minimizes the need for chiller setpoint changes when testing to the ASTM 5470 sample average-temperature of 50°C some level of inconvenience will be experienced under manual chiller control depending on the test specifications selected. .

WinTIM software supports serial-equipped Thermo-Fischer and Lytron chillers with serial command tables that are embedded in WinTIM software. The following chillers all provide these serial control features and are recommended:

Thermo-Fischer Scientific Chillers
▪ 100-115VAC, 50/60 Hz TF9 B A 100-115/GL P1 IPR DCOM
▪ 200-230VAC, 50/60 Hz TF9 B A 200-230/GL P1 IPR DCOM

Lytron Chillers
▪ 115VAC, 60 Hz RC006-GO3-BE-3
▪ 230VAC, 50 Hz RC006-HO3-BE-3
▪ 230VAC, 60 Hz RC006-JO3-BE-3

Ordering Information
These chillers can be ordered directly from Thermo-Fischer Scientific or Lytron Inc.

 
 