The Phase 10/11 Instrument Calibrator is used to perform NIST traceable instrument calibrations on the Phase 10 and Phase 11 Thermal Analyzers, typically at 6 month intervals. The Calibrator facilitates a simple, ten-minute procedure for calibrating a thermal analyzer to within optimal measurement specifications. This precise instrument calibration ensures that the test data generated by the Thermal Analyzer will be of the highest quality and accuracy. The calibrated parameters include:

Heating-Power:
All voltage and current measurement

Temperature Sensing:
Functional check of all temperature-sense channels
Accuracy test of all channels at all sense current levels

Heating Power Control and Delivery:
Complete scan of heating power delivery ports
Test and alignment of all power control regulators

The Phase 10/11 Instrument Calibrator is a robust, precision reference unit that interfaces with the Thermal Analyzer heating power and junction temperature sensing connectors. This small unit can be conveniently returned to the Analysis Tech factory every two years for re-certification to NIST traceable calibration standards.

Electrical Specifications

Voltage Sensing Accuracy (0 to 20V) +/- 5 mV
Low Current Sensing Accuracy (0 to 2A) +/- 1 mA
High Current Sensing Accuracy (0 to 20A) +/- 5 mA
Sense Current Measurement Accuracy +/- 0.24%

Mechanical Specifications
Dimensions 3" H x 9" W x 6" D
Weight 3 lbs

Ordering Information
Part Number: Instrument Calibrator. Standard delivery 3 weeks.
Ex Works Wakefield, MA USA
 
 