All Thermal Analyzers are available with either 2, 10 or 20 amp maximum heating current capacity, indicated by placing a "-2", a "-10" or a "-20" after the model number. The following specifications are labeled where differences exist between the 2, 10 and 20 amp units; otherwise, specifications apply to all versions. AC power requirements are assumed to be 120 VAC, 50/60Hz unless otherwise specified at the time ordering. Options are available to extend the maximum heating voltages beyond 50 V and can be discussed with Analysis Tech engineers. For higher power testing, the Power Booster can extend the current capacity to 400A and heating power delivery to 4000W.

Heating current, high-range measurement accuracy:
  2A system: ±4mA 10A system: ±20mA 20A system: ±40mA

Heating current, low-range measurement accuracy:
  2A system: ±1mA 10A system: ±5mA 20A system: ±10mA

Heating voltage measurement accuracy:

  ±0.2% of reading ±0.025% of full scale (typical 50V full scale)

Thermocouple measurement accuracy (type T standard):
  ±0.1°C typical, ±0.3°C maximum  

Junction temperature measurement accuracy:
  ±0.1°C typical  

Junction Temperature measurement delay: 1 microsecond minimum
AC Power Supply: 120VAC, 3A, 50/60Hz (standard unless otherwise specified)
Maximum Heating Supply Voltage: 50 volts (standard) ranges to 300V available
Maximum Heating Supply Current: 2, 10 or 20 amps without Power Booster
Tj Sense Current Selections: 1mA, 5mA, 10mA, 20mA, 50mA, and adjustable 0.1 - 50mA

Phase 11 Chassis Dimensions : 17" x 17" x 7.5" high, 25 lbs

 
 