All Thermal Analyzers are available with either 2, 10 or 20 amp maximum heating current capacity, indicated by placing a "-2", a "-10" or a "-20" after the model number. The following specifications are labeled where differences exist between the 2, 10 and 20 amp units; otherwise, specifications apply to all versions. AC power requirements are assumed to be 120 VAC, 50/60Hz unless otherwise specified at the time ordering. Options are available to extend the maximum heating voltages beyond 50 V and can be discussed with Analysis Tech engineers. For higher power testing, the Power Booster can extend the current capacity to 400A and heating power delivery to 4000W.

| Heating current, high-range measurement accuracy: |
| |
2A system: ±4mA |
10A system: ±20mA |
20A system: ±40mA |
Heating current, low-range measurement accuracy: |
| |
2A system: ±1mA |
10A system: ±5mA |
20A system: ±10mA |
Heating voltage measurement accuracy:
|
| |
±0.2% of reading ±0.025% of full scale (typical 50V full scale) |
| Thermocouple measurement accuracy (type T standard): |
| |
±0.1°C typical, ±0.3°C maximum |
|
| Junction temperature measurement accuracy: |
| |
±0.1°C typical |
|
Junction Temperature measurement delay: 1 microsecond minimum
AC Power Supply: 120VAC, 3A, 50/60Hz (standard unless otherwise specified)
Maximum Heating Supply Voltage: 50 volts (standard) ranges to 300V available
Maximum Heating Supply Current: 2, 10 or 20 amps without Power Booster
Tj Sense Current Selections: 1mA, 5mA, 10mA, 20mA, 50mA, and adjustable 0.1 - 50mA
Phase 11 Chassis Dimensions : 17" x 17" x 7.5" high, 25 lbs
|
|